Sign In | Join Free | My himfr.com
Home > Patient Monitor Parts >

PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now
    Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
     
    Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers
    • Buy cheap PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 from wholesalers

    PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

    Ask Lasest Price
    Brand Name : PHiliph
    Model Number : E131992 J09L078118
    Payment Terms : T/T(wire transfer), Western Union, Paypal
    Delivery Time : 2-3 working days
    MOQ : One piece
    • Product Details
    • Company Profile

    PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118

    PHiliphh MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118


    Touchscreen Calibration


    To access the touchscreen calibration screen:


    1 Enter service mode


    2 Select Main Setup


    3 Select Hardware


    4 Select Touch Calibration


    Perform the calibration procedure without turning off the monitor. If the screen turns off after touching the first point, the touch screen will be disabled until touch calibration is performed again.

    If you calibrate the touchscreen incorrectly by selecting the wrong location, you will need to re-enter calibration mode using a different input device. If you have a support device, you can choose to reset touch calibration to the default. This creates an approximate calibration that allows you to return to the calibration menu from the touch screen.


    Integrated Touch Display not functioning


    1.Touch Screen not functioning

    Possible Causes of FailureFailure Isolation and Remedy
    The touchscreen function is temporarily disabled

    Check if the touch screen function is temporarily disabled (lock symbol on the home button). If yes, press and hold the main button

    The on-screen button reactivates the touchscreen.

    The touchscreen function is permanently disabledIn Service Mode, select Main Settings -> User Interface and change "Enable Touch" to "Yes".
    Tablet adapter not connectedCheck the connection between the tablet adapter and the motherboard
    The touch controller is not connectedCheck the connection between the touch controller and the tablet adapter
    Touch screen cable not connectedCheck the connection between touch control panel and touch panel
    Defective touch control board or touch sensor

    Touch control panel and touch sensor replacement

    Note: Linearization data must be uploaded and recalibrated later

    Touch control panel replacement

    the motherboard is brokenReplace motherboard

    2.Invalid touch position


    Possible Causes of FailureFailure Isolation and Remedy
    the touch is not calibrated

    To calibrate by touch:

    1. Enter the main settings menu

    2. Select hardware

    3. Select the touch adapter settings

    4. Select Calibrate


    Disable/enable touch actions

    There are two ways to disable/enable the touchscreen functionality:

    1. Press and hold the home button to temporarily disable the touchscreen functionality. A padlock symbol appears on the key. Press and hold the home button again to reactivate the touchscreen function.

    2. To permanently disable the touchscreen:

    A. Enter service mode.

    B. Select main settings

    C. Select User Interface

    D. Change Enable Touch to no.

    Set Enable Touchscreen to Yes to re-enable touchscreen functionality


    Removing the Touchscreen


    1 Remove the front and back of the monitor as above.

    2 Remove the touch controller board, press the connector safety latch, and disconnect the touch sensor cable.

    Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.

    Assembly Note: First connect the touch sensor cable to the touch controller board, then reinstall the touchpad.


    3. Remove the two screws securing the power button and LED board and remove the board. Plates and screws are part of the flex panel adapter cable.

    Note The screws that secure this plate are self-tapping screws. When reinstalling, first turn it in the opposite direction.

    4. Remove the 8 screws, loosen the 2 clips on the monitor frame and carefully remove the monitor assembly.

    NOTE: It is recommended that this procedure be performed in a dust-free environment. The touchscreen display assembly can be difficult to remove due to the gaskets (tape) holding the two together.

    5 Remove the touch screen (sensor) from the frame.

    Note This operation can be difficult due to the seams (bands).

    Assembly Note: When viewing the touch screen from the rear, the touch sensor cable must be facing to the right when placing the touch screen back into the monitor housing.

    Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.


    If you're interested in PHiliphh MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 ,feel free to contact us .Thank you .

    Quality PHiliph MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: China World Technology Medical Equipment Service Group
    *Subject:
    *Message:
    Characters Remaining: (0/3000)